Biosafety Cabinets & Laminar Flow Hoods: Ensuring Contamination Control in Critical Workflows
Key Applications
1. Biosafety Cabinets (Class II A2):
-
Virology: Handling SARS-CoV-2 cultures with inward airflow (0.38–0.51 m/s) and HEPA-filtered exhaust.
-
Cell Therapy: Processing CAR-T cells under aseptic conditions (EU GMP Annex 1 compliance).
-
Mycobacteriology: Containing M. tuberculosis via negative-pressure plenums and UV decontamination cycles.
2. Laminar Flow Hoods (Vertical/Horizontal):
-
Pharmaceutical Compounding: Preparing IV admixtures in USP <797>-compliant environments.
-
Electronics Assembly: Preventing particulate contamination during microchip manufacturing.
-
Plant Tissue Culture: Transferring explants under unidirectional airflow (0.45 m/s ±20%).
Technical Specifications & Innovations
Biosafety Cabinet (Class II A2)
-
Airflow System:
-
30% recirculated + 70% exhausted HEPA (H14) airflow
-
Inflow velocity: 0.53 m/s ±0.025 (EN 12469:2000 validated)
-
-
Filtration:
-
Dual ULPA filters (99.9995% @ 0.12μm) for exhaust
-
Optional germicidal UV-C (254 nm) with 25 μW/cm² irradiance
-
-
Ergonomics:
-
Anti-glide sash with 8–18 inch working height
-
Foot-operated electrical outlets (IEC 61010-1 compliant)
-
Laminar Flow Hood (Vertical)
-
Airflow Design:
-
Unidirectional vertical flow (0.3–0.5 m/s) over 0.24 m² workspace
-
Pre-filter (G4) + HEPA (H13) filtration (99.95% @ 0.3μm)
-
-
Construction:
-
304 stainless steel interior (Ra ≤0.8 μm roughness)
-
Seamless curved corners for CIP (Clean-in-Place) compliance
-
-
Monitoring:
-
Magnehelic gauge (±5% accuracy) for filter load tracking
-
Real-time particle counter (ISO 14644-1 Class 5)
-